Overview CANNON®
through the typical range of spindle speeds for the sample type or expected viscosity. This
eliminates the need to run multiple tubes of a sample. Once the viscosity measurements
finish, the TESC System returns the sample to room temperature, nominally 25 °C.
Afterwards, the operator can review the data and enhance the digital record with notes and
additional information. The TESC System comes with a set of thermal conditioning programs
for all of the common specification test temperatures as well as certain OEM measurement
and report requirements.
Table 1 — TESC Thermal conditioning programs
Fluid
Preheat Temperature
Test Temperature
Low viscosity (ATF, hydraulic fluid) 50 °C
10 °C
20 °C
30 °C
40 °C
High viscosity (gear oil) 90 °C
12 °C
26 °C
40 °C
Related documents
Table 2 — Related documents
Title
Reference Number
Standard Test Method for Low-Temperature Viscosity of
Lubricants Measured by Brookfield® Viscometer
ASTM D2983
Automotive Gear Lubricant Viscosity Classification
SAE J306