NOTE
International shipments may be subject to special shipping regulations.
ASTM D 445 — Checking the ice point
Frequency
To achieve an accuracy of ± 0.02°C for calibrated kinematic viscosity
thermometers, a check at the ice point must be made. New thermometers
should be checked monthly for the first six months, then once every six
months.